l |
Prospective
Gating: 0.4 sec gantry speed.
|
|
|
Machine counts
5 cycles, calculates R-R
|
|
|
interval,
takes 0.25 sec scan, ending prior
|
|
|
to beginning
of next R wave. Requires
|
|
|
heart rate
< 80 bpm. Table moves during
|
|
|
next R-R
interval. Each scan covers 4
|
|
|
images @ 3mm
thickness.
|
|
|
l |
Retrospective
gating = gated
|
|
|
reconstruction. Helical acquisition with
|
|
|
ECG over raw
data. When ECG is at a given
|
|
point, take
data at that time to make image.
|
|