16 slice scanner
l Prospective Gating: 0.4 sec gantry speed.
Machine counts 5 cycles, calculates R-R
interval, takes 0.25 sec scan, ending prior
to beginning of next R wave.  Requires
heart rate < 80 bpm.  Table moves during
next R-R interval.  Each scan covers 4
images @ 3mm thickness.
l Retrospective gating =  gated
reconstruction.  Helical acquisition with
ECG over raw data.  When ECG is at a given
point, take data at that time to make image.